| HTC 08/15 with lift door | HTC 03/14 with fold-down door and gas supply system |
Furnace chamber with high-quality fibre material and SiC heating rods in the sides of the furnace
Saggar
Spacer
Over-temperature limit controller
These powerful laboratory muffle furnaces are available for temperatures up to 1400 °C, 1500 °C, or 1600 °C. The durability of the SiC rods in periodic use, in combination with their high heating speed, make these furnaces to all-rounders in the laboratory. Heating times of 40 minutes to 1400 °C can be achieved, depending on the furnace model and the conditions of use.
- Tmax 1400 °C, 1500 °C, or 1600 °C
- Easy replacement of heating rods
- High-quality fibre material, selected for the working temperature
- Casing made of sheets of textured stainless steel (non-rusting design)
- Double-walled casing for low external temperatures and high stability
- Optional fold-down door which can be used as work platform or lift door with hot surface facing away from the operator
- Adjustable air intake opening in the furnace door, exhaust air opening in the back wall
- Switching system with solid-state-relays, power tuned to the SiC rods
Additional Equipment
- Over-temperature limit controller with adjustable cutout temperature for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the oven and load
- Saggar for charging in multiple layers
- Spacer recommended to be placed under the bottom saggar
| Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Power | Electrical | Weight | Minutes | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| °C | w | d | h | in L | W | D | H | kW | connection* | in kg | to Tmax | |
| HTC 03/14 | 1400 | 120 | 210 | 120 | 3 | 400 | 535 | 530 | 9,0 | 3-phase¹ | 30 | 40 |
| HTC 08/14 | 1400 | 170 | 290 | 170 | 8 | 450 | 620 | 570 | 13,0 | 3-phase | 40 | 40 |
| HTC 03/15 | 1500 | 120 | 210 | 120 | 3 | 400 | 535 | 530 | 9,0 | 3-phase¹ | 30 | 50 |
| HTC 08/15 | 1500 | 170 | 290 | 170 | 8 | 450 | 620 | 570 | 13,0 | 3-phase | 40 | 50 |
| HTC 03/16 | 1600 | 120 | 210 | 120 | 3 | 400 | 535 | 530 | 9,0 | 3-phase¹ | 30 | 60 |
| HTC 08/16 | 1600 | 170 | 290 | 170 | 8 | 450 | 620 | 570 | 13,0 | 3-phase | 40 | 60 |
| ¹Heating only between two phases *Please see page 50 for more information about mains voltage | ||||||||||||
![]() |
![]() |
|
| LHT 08/17 | LHT 02/17 |
Saggar
Spacer
Over-temperature limit controller
Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages. The first-class workmanship using high-quality materials, combined with ease of operation, make these furnaces all-rounders in research and the laboratory. These furnaces are also perfectly suited for the sintering of technical ceramics, such as zirconium oxide dental bridges.
- Tmax 1600 °C, 1750 °C, or 1800 °C
- High-quality molybdenum disilicide heating elements
- Furnace chamber lined with first-class, durable fibre material
- Casing made of sheets of textured stainless steel (non-rusting design)
- Double-walled casing with additional cooling using fans, for low exterior temperature
- Furnace sizes of 2, 4, or 8 liters
- With lift door, whereby the hot side is away from the operator
- Adjustable air inlet integrated in door
- Exhaust air opening in the roof
- Type B thermocouple
- Switching system with phase-angle firing thyristors (SCRs)
Additional Equipment
- Over-temperature limit controller with adjustable cutout temperature for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the oven and load
- Saggar for charging in multiple layers
- Spacer recommended to be placed under the bottom saggar
- Protective gas connector in rear wall of furnace
- Manual or automatic gas supply system
| Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Power | Electrical | Weight | Minutes | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| °C | w | d | h | in L | W | D | H | kW | connection* | in kg | to Tmax | |
| LHT 02/16 | 1600 | 90 | 150 | 150 | 2 | 470 | 700 | 750+350 | 3,0 | single-phase | 75 | 30 |
| LHT 04/16 | 1600 | 150 | 150 | 150 | 4 | 470 | 700 | 750+350 | 5,2 | 3-phase¹ | 85 | 25 |
| LHT 08/16 | 1600 | 150 | 300 | 150 | 8 | 470 | 850 | 750+350 | 8,0 | 3-phase¹ | 100 | 25 |
| LHT 02/17 | 1750 | 90 | 150 | 150 | 2 | 470 | 700 | 750+350 | 3,0 | single-phase | 75 | 60 |
| LHT 04/17 | 1750 | 150 | 150 | 150 | 4 | 470 | 700 | 750+350 | 5,2 | 3-phase¹ | 85 | 40 |
| LHT 08/17 | 1750 | 150 | 300 | 150 | 8 | 470 | 850 | 750+350 | 8,0 | 3-phase¹ | 100 | 40 |
| LHT 02/18 | 1800 | 90 | 150 | 150 | 2 | 470 | 700 | 750+350 | 3,6 | single-phase | 75 | 75 |
| LHT 04/18 | 1800 | 150 | 150 | 150 | 4 | 470 | 700 | 750+350 | 5,2 | 3-phase¹ | 85 | 60 |
| LHT 08/18 | 1800 | 150 | 300 | 150 | 8 | 470 | 850 | 750+350 | 9,0 | 3-phase¹ | 100 | 60 |
| ¹Heating only between two phases *Please see page 50 for more information about mains voltage | ||||||||||||
![]() |
![]() |
|
| LHT 02/17 LB closed, with over-temperature limit controller | LHT 02/16 LB with saggar and over-temperature limit controller |
Plasma ceramic retort for protecting the heating elements and to avoid charge contimination [in the LHT 02/16 LBR]
Saggar
Due to the electrically driven lift bottom, the LHT/LB series high-temperature furnaces can be charged very handily. With their all-round heating, these furnaces achieve outstanding temperature uniformity. In the LHT 02/16 LBR model, the furnace chamber is separated from the heating elements due to the use of a tubular plasma ceramic retort placed vertically in the furnace. This design prevents a certain degree of contamination of both the load and the heating elements. It is also recommended for gassing processes.
- Tmax 1600 °C or 1700 °C
- High-quality molybdenum disilicide heating elements
- Furnace chamber lined with first-class, durable fiber materials
- Outstanding temperature uniformity due to all-round furnace chamber heating
- Tubular plasma ceramic retort for the LBR version to prevent a certain level of contamination and to improve temperature uniformity
- Furnace chamber with a volume of 2 liters, table with large footprint
- Precise, electric spindle drive with push button operation
- Casing made of sheets of textured stainless steel (non-rusting design)
- Adjustable air inlet through the floor
- Exhaust air vent in the roof
- Type B thermocouple
- Switchgear with thyristor
Additional Equipment
- Over-temperature limit controller with adjustable cutout temperature for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the furnace and load
- Saggar for charging in up to three layers
- Protective gas connection
- Manual or automatic gas supply system, particularly effective in the LBR version
| Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Power | Electrical | Weight | Minutes | |||
|---|---|---|---|---|---|---|---|---|---|---|---|
| °C | Ø | h | in L | W | D | H | kW | connection* | in kg | to Tmax | |
| LHT 02/16 LB | 1600 | Ø 120 | 130 | 2 | 540 | 610 | 740 | 3,0 | single-phase | 65 | approx. 85 |
| LHT 02/16 LBR | 1600 | Ø 120 | 130 | 2 | 540 | 610 | 740 | 3,0 | single-phase | 65 | approx. 85 |
| LHT 02/17 LB | 1700 | Ø 120 | 130 | 2 | 540 | 610 | 740 | 3,0 | single-phase | 65 | approx. 85 |
| *Please see page 50 for more information about mains voltage | |||||||||||
![]() |
![]() |
| HT 16/16 | HT 32/17 with automatic vapour vent flap and fresh air fan with dimmer |
Gas supply panel for nonflammable protective gas with shutoff valve and flow meter with regulator valve, piped and ready to connect
Automatic vent flap
Due to their solid construction and compact stand-alone design, these high-temperature furnaces are perfect for processes in the laboratory where the highest precision is needed. Oustanding temperature uniformity and practical details set unbeatable quality benchmarks. For configuration for your processes, these furnaces can be extended with extras from our extensive option list.
- Tmax 1600 °C, 1750 °C, or 1800 °C
- High-quality molybdenum disilicide (MoSi2) heating elements
- Furnace chamber lined with first-class, durable fibre material
- Parallel swinging doors, chain-guided, allowing safe opening and closing without destruction of the fibre insulation. Positive closing of the door with a cam latch
- Type B thermocouple
- Furnace sizes from 4 to 450 liters, special sizes on request
- Over-temperature limit controller as standard equipment for protection of material
Additional Equipment
- Floor reinforcement for support of heavy loads
- Exhaust air flap controlled manually or automatically for better ventilation of furnace chamber
- Fan for better ventilation of combustion chamber and for fast cooling of the furnace
- Protective gas connector and seal of furnace casing to allow purging of furnace with protective gasses
- Manual or automatic gas supply system
| Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Power | Electrical | Weight | Minutes | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| °C | w | d | h | in L | W | D | H | kW | connection* | in kg | to Tmax | |
| HT 04/16 | 1600 | 150 | 150 | 150 | 4 | 610 | 470 | 1400 | 5,2 | 3-phase¹ | 150 | 25 |
| HT 08/16 | 1600 | 150 | 300 | 150 | 8 | 610 | 610 | 1400 | 8,0 | 3-phase¹ | 200 | 25 |
| HT 16/16 | 1600 | 200 | 300 | 260 | 16 | 710 | 650 | 1500 | 12,0 | 3-phase¹ | 270 | 25 |
| HT 04/17 | 1750 | 150 | 150 | 150 | 4 | 610 | 470 | 1400 | 5,2 | 3-phase¹ | 150 | 40 |
| HT 08/17 | 1750 | 150 | 300 | 150 | 8 | 610 | 610 | 1400 | 8,0 | 3-phase¹ | 200 | 40 |
| HT 16/17 | 1750 | 200 | 300 | 260 | 16 | 710 | 650 | 1500 | 12,0 | 3-phase¹ | 270 | 40 |
| HT 04/18 | 1800 | 150 | 150 | 150 | 4 | 610 | 470 | 1400 | 5,2 | 3-phase¹ | 150 | 40 |
| HT 08/18 | 1800 | 150 | 300 | 150 | 8 | 610 | 610 | 1400 | 9,0 | 3-phase¹ | 200 | 40 |
| HT 16/18 | 1800 | 200 | 300 | 260 | 16 | 710 | 650 | 1500 | 12,0 | 3-phase¹ | 270 | 40 |
| ¹Heating only between two phases *Please see page 50 for more information about mains voltage | ||||||||||||
![]() |
![]() |
| HFL 16/16 | HFL 295/13 with lift door and transformer in stand, with customer-specific design |
Protective screen in front of heating elements for protection against mechanical damage
The HFL 16/16 - HFL 160/17 series is particularly characterized by their special cladding with light refractory bricks. This insulation is necessary when aggressive gasses (such as vapourized glass) or acids may be produced in the process.
- Tmax 1600 °C or 1700 °C
- High-quality molybdenum disilicide (MoSi2) heating elements
- Insulation with light refractory bricks and special backup insulation
- Type B thermocouple
- Furnace sizes of 16 to 160 liters
- For the release of vapours, a 30 mm large exhaust hole is integrated into the roof of the furnace
- Over-temperature limit controller for protection of material
Additional Equipment
- Exhaust air flap controlled manually or automatically for better ventilation of furnace chamber
- Fan for better ventilation of combustion chamber and for fast cooling of the furnace
- Protective gas connector and seal of furnace casing to allow purging of furnace with protective gasses
- Manual or automatic gas supply system
| Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Power | Electrical | Weight | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| °C | w | d | h | in L | W | D | H | kW | connection* | in kg | |
| HFL 16/16 | 1600 | 200 | 300 | 260 | 16 | 770 | 830 | 1550 | 12 | 3-phase¹ | 500 |
| HFL 40/16 | 1600 | 300 | 350 | 350 | 40 | 880 | 880 | 1710 | 12 | 3-phase | 660 |
| HFL 64/16 | 1600 | 400 | 400 | 400 | 64 | 980 | 930 | 1830 | 18 | 3-phase | 880 |
| HFL 160/16 | 1600 | 500 | 550 | 550 | 160 | 1090 | 1080 | 2030 | 21 | 3-phase | 1140 |
| HFL 16/17 | 1700 | 200 | 300 | 260 | 16 | 770 | 830 | 1550 | 12 | 3-phase¹ | 530 |
| HFL 40/17 | 1700 | 300 | 350 | 350 | 40 | 880 | 880 | 1710 | 12 | 3-phase | 690 |
| HFL 64/17 | 1700 | 400 | 400 | 400 | 64 | 980 | 930 | 1830 | 18 | 3-phase | 920 |
| HFL 160/17 | 1700 | 500 | 550 | 550 | 160 | 1090 | 1080 | 2030 | 21 | 3-phase | 1190 |
| ¹Heating only between two phases *Please see page 50 for more information about mains voltage | |||||||||||