Muffle Furnace incl. Scale and Software for Determination of Combustion Loss

Delivery includes base, ceramic plunger with base plate in the furnace lining, precision scale and software package.

Process Control

Function and equipment

L9_11_SW_0.png
Waage_fmt.jpg
TWB_0.png
Grafische_Darstellung.png

This weighing furnace with integrated precision scale and software, was designed especially for combustion loss determination in the laboratory. The determination of combustion loss is necessary, for instance, when analyzing sludges and household garbage, and is also used in a variety of other processes for the evaluation of results. The difference between the charged total mass and the combustion residue is the combustion loss. During the process, the software included records both the temperature and the weight loss.

Standard Equipment

  • Like muffle furnaces L(T), except:
  • Delivery includes base, ceramic plunger with base plate in the furnace lining, precision scale and software package
  • 4 scales available for different maximum weights and scaling ranges
  • Process control and documentation for temperature and combustion loss via VCD software package for monitoring, documentation and control
  • Controller B510 (5 programs with each 4 segments)

Additional equipment

Additional Equipment

  • Chimney, chimney with fan or catalytic converter
  • Over-temperature limiter with adjustable cutout temperature as temperature limiter to protect the furnace and load
  • Port for thermocouple in the rear wall or in the furnace door
  • Process control and documentation via VCD software package for monitoring, documentation and control

Technical specifications

Model Tmax Inner dimensions in mm Volume Outer dimensions2 in mm Connected Electrical Weight Heating time
  in °C1 w d h in l W D H load in kW connection* in kg in min4
L(T) 9/11/SW 1100 230 240 170 9 415 455 740+2403 3.0 1-phase 50 65
L(T) 9/12/SW 1200 230 240 170 9 415 455 740+2403 3.0 1-phase 50 75
1Recommended working temperature for processes with longer dwell times is 1000 °C (L 9/11) rsp. 1100 °C (L 9/12)*Please see page 75 for more information about supply voltage
2External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.
3Including opened lift door (Model LT ..)
4Heating time of the empty and closed furnace up to Tmax −100 K (connected to 230 V 1/N/PE)

 

Scale Readability Maximum weighing range Weight of plunger Calibration value Minimum load
type in g in g in g in g in g
EW-2200 0.01 2200 incl. plunger 850 0.1 0.5
EW-4200 0.01 4200 incl. plunger 850 0.1 0.5
EW-6200 0.01 6200 incl. plunger 850 - 1.0
EW-12000 0.10 12000 incl. plunger 850 1.0 5.0

Request for quotation